ISL OpenIR  > 青海盐湖研究所知识仓储
A comparison of loading reagent for boron isotopic measurement bynegative thermal ionization mass spectrometry
Liu, WG; Peng, ZC; Xiao, YK; Zhang, ZF
2002
发表期刊CHINESE JOURNAL OF ANALYTICAL CHEMISTRY
ISSN0253-3820
卷号30期号:3页码:372-375
摘要The isobaric (CNO- ions) interference involved in the loading reagents has not been noted for boron isotopic measurement by nagetive thermal ionization mass spectrometry (NTIMS). A comparison experiment of different loading reagents was investigated. The result revealed that high isobaric interference ion (CNO-) for boron isotopic measurement was observed in the boron-free seawater and the nitrate. Further, the CNO- was not only from the organic material, but also from nitrate in loading reagent in NTIMS. The isobaric interference in loading reagents may be a reason resulting to result in low boron isotopic ratios by NTIMS. So monitoring the mass 43 ion intensity and 43/42 ratio of blank are necessary for the boron isotopic measurement by NTIMS. The mixed solution of the MgCl2 and NaOH is a suitable loading reagent for boron isotopic measurement by NTIMS. The isobaric (CNO- ions) interference involved in the loading reagents has not been noted for boron isotopic measurement by nagetive thermal ionization mass spectrometry (NTIMS). A comparison experiment of different loading reagents was investigated. The result revealed that high isobaric interference ion (CNO-) for boron isotopic measurement was observed in the boron-free seawater and the nitrate. Further, the CNO- was not only from the organic material, but also from nitrate in loading reagent in NTIMS. The isobaric interference in loading reagents may be a reason resulting to result in low boron isotopic ratios by NTIMS. So monitoring the mass 43 ion intensity and 43/42 ratio of blank are necessary for the boron isotopic measurement by NTIMS. The mixed solution of the MgCl2 and NaOH is a suitable loading reagent for boron isotopic measurement by NTIMS.
文献类型期刊论文
条目标识符http://ir.isl.ac.cn/handle/363002/2344
专题青海盐湖研究所知识仓储
中国科学院青海盐湖研究所
推荐引用方式
GB/T 7714
Liu, WG,Peng, ZC,Xiao, YK,et al. A comparison of loading reagent for boron isotopic measurement bynegative thermal ionization mass spectrometry[J]. CHINESE JOURNAL OF ANALYTICAL CHEMISTRY,2002,30(3):372-375.
APA Liu, WG,Peng, ZC,Xiao, YK,&Zhang, ZF.(2002).A comparison of loading reagent for boron isotopic measurement bynegative thermal ionization mass spectrometry.CHINESE JOURNAL OF ANALYTICAL CHEMISTRY,30(3),372-375.
MLA Liu, WG,et al."A comparison of loading reagent for boron isotopic measurement bynegative thermal ionization mass spectrometry".CHINESE JOURNAL OF ANALYTICAL CHEMISTRY 30.3(2002):372-375.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Liu, WG]的文章
[Peng, ZC]的文章
[Xiao, YK]的文章
百度学术
百度学术中相似的文章
[Liu, WG]的文章
[Peng, ZC]的文章
[Xiao, YK]的文章
必应学术
必应学术中相似的文章
[Liu, WG]的文章
[Peng, ZC]的文章
[Xiao, YK]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。